Export Electrode Burst Files from AxIS Navigator 3.6.2 software

The following vignette describes the workflow used to analyse Axion Biosystems MEA recordings and export the associated electrode burst csv files for use in the MEAanalysis package.

  1. Load your Axion Biosystems MEA recording into the AxIS Navigator 3.6.2 software (click file, click open recordings, navigate to your recording and click open).

  2. Right click the chosen recording in the ‘Streams’ panel and click configuration. Select the spontaneous and viability configuration.

  3. Right click the Spike Detector and selective an Adaptive Threshold Crossing of 6x standard deviation.

  4. Right click the Burst Detector (ISI) and set the Maximum Inter-Spike Interval to 100 mS and the Minimum Number of Spikes to 5.